Abstract:In the real environment, there is a crosssensitivity problem between fiber grating strain and temperature measurement. This paper proposes a simultaneous measurement method of two parameters using chirp fiber grating sensor. By sealing chirp fiber bragg grating (CFBG) on the equalstrength beam, the difference between the sensitivity of the center wavelength and the bandwidth of the CFBG reflection spectrum to temperature and strain is used to form a coefficient decoupling matrix, which can achieve simultaneous measurement of the corresponding change and temperature. At the room temperature, the sensitivities of CFBG center wavelength and bandwidth with strain are 079 and 138 pm/με. The corresponding linearity values are 0998 8 and 0999 3. Among the temperature range of -20~60℃, the sensitivities of the central wavelength and bandwidth of CFBG with temperature are 2274 and 2397 pm/℃. The linearity values are 0999 8 and 0997 0. Experimental results show that the simultaneous measurement of strain and temperature can be achieved using one single CFBG.