Design of a flexible non-contact ECG array electrodes and effect analysis on pressure
Author:
Affiliation:

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    非接触式心电传感技术在长期睡眠心电监测领域展现了优异的应用前景,其中阵列式非接触心电传感技术取得了较 好的发展,然而,该技术所用电极多为刚性的印制电路板(PCB),舒适性较差。 同时,人体压力分布对该技术所获心电信号质量 的影响规律尚未可知。 针对这些不足,本研究设计了一种基于柔性印刷电路(FPC)的阵列式非接触睡眠心电监测系统。 通过 系统验证,证明了该系统所获取心电信号一致性较高且平均信噪比>38 dB。 随后通过人体试验分析了人体压力对心电信号质 量的影响。 结果表明,当心电检测输入端所受压力<9 mmHg 时,获得低信噪比心电信号的概率 P 高于 54. 17% ;反之,则 P 低于 5. 94% ,证明人体压力是影响电容耦合心电信号质量的关键因素之一。

    Abstract:

    The non-contact electrocardiogram (ECG) sensing technology has a promising application in the field of long-term sleeping ECG monitoring. The non-contact ECG sensing technology based on array electrodes has made good progress. However, the current noncontact ECG sensing electrodes based on array electrodes are mostly manufactured by using the hard printed circuit board technology, which has low comfort. Meanwhile, the relationship between the human body pressure distribution and the signal quality of non-contact ECG sensing based on array electrodes is not studied in this aspect based on our best knowledge. To address those issues, this study develops an arrayed non-contact ECG sensing electrode based on flexible printed circuit (FPC). The experimental results show that the ECG signals obtained from the proposed system are high, and the averaged signal-to-noise ratio is larger than 38 dB. Moreover, the effect of pressure distribution of human back on the ECG signal quality is analyzed by conducting the human trials. The results demonstrate that the obtained ECG signal quality is relatively low when the electrode unit with human body pressure is lower than 9 mmHg used as one of the positive or negative ECG signal input units. The probability for obtaining ECG signals with lower quality is higher than 54. 17% . When the unit pressure is higher than 9 mmHg, the probability for obtaining ECG signals with lower quality is lower than 5. 94% . It proves that the unit pressure is one of the key factors that can affect the quality of capacitive ECG signals.

    参考文献
    相似文献
    引证文献
引用本文

吴 刚,冯宝亮,刘 箴,左国坤,施长城.柔性非接触心电阵列电极设计与压力影响分析[J].仪器仪表学报,2023,44(12):150-160

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:
  • 最后修改日期:
  • 录用日期:
  • 在线发布日期: 2024-02-27
  • 出版日期: