臭氧标准参考光度计的测量不确定度评定
DOI:
作者:
作者单位:

环境保护部标准样品研究所 国家环境保护污染物计量和标准样品研究重点实验室 北京 100029

作者简介:

通讯作者:

中图分类号:

TH744.12X831

基金项目:

国家环境保护公益性行业科研专项(201409011)、国家环境保护标准制修订项目(201461)、环保部环境发展中心自主选题科技(ZZ201604)项目资助


Measurement uncertainty evaluation for ozone standard reference photometer
Author:
Affiliation:

Institute for Environmental Reference Materials of Ministry of Environmental Protection,State Environmental Protection Key Laboratory of Pollutants Metrology and Reference Materials Research,Beijing 100029,China

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    为建立臭氧标准参考光度计(SRP)的计量溯源性,环保部标准样品研究所的SRP 48与美国国家标准与技术研究院的SRP 2进行了比对验证。对SRP测量不确定度的评定方法进行了研究,分析了测量不确定度的主要来源。以SRP 48为例,对影响臭氧浓度测量的不确定度分量光程长度Lopt、压力P、温度T、双光程透光率的比值D和吸收系数σ等因素进行计算,建立数学模型并合成得到标准不确定度。在0~500 nmol/mol浓度范围内,SRP 48和SRP 2之间具有很好的可比性,确立SRP 48与SRP 2的比对关系式为xSRP48=[(0.998 96(xSRP2)+0.025] nmol/mol,在95%的置信区间内(k=2时),SRP 48臭氧浓度测量的扩展不确定度为2×((0.28)2+(1.10×10-2×x)2) nmol/mol。

    Abstract:

    The comparison and validation of ozone Standard Reference Photometer (SRP 48) by Institute for Environmental Reference Materials of MEP(IERM of MEP) and SRP 2 by National Institute of Standards and Technology(NIST) are performed in order to establish the metrological traceability. The evaluating the measurement uncertainty of SRP is focused and the main sources of measurement uncertainty are identified. Take SRP 48 for example, the measurement uncertainty components such as the optical path length (Lopt), the pressure (P), the temperature (T), the product of transmittances ratio of the two cells (D) and the crosssection (σ) are calculated, respectively. A measurement mathematical model is established and the standard uncertainty is combined. The results of the comparison show that SRP 48 and SRP 2 are comparable over an ozone mole fraction range of 0~500 nmol/mol. The relationship between SRP 48 and SRP 2 is xSRP48= [(0.998 96(xSRP2)+0.025] nmol/mol. The expanded uncertainty of SRP 48 ozone concentration measurement is 2×((0.28)2+(1.10×10-2×x)2) nmol/mol at 95% confidence interval using a coverage factor k=2.

    参考文献
    相似文献
    引证文献
引用本文

王帅斌,李宁,樊强,田文,王倩.臭氧标准参考光度计的测量不确定度评定[J].仪器仪表学报,2017,38(12):3003-3012

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:
  • 最后修改日期:
  • 录用日期:
  • 在线发布日期: 2018-01-17
  • 出版日期: