Abstract:With the development of force metrology within superlarge range,force transfer system with a superlarge range has become a hotspot in the field of force metrology recent years. An uncertainty calculation model of the buildup system with a superlarge rangeconsidering the influence of the buildup structure was developed. A 60 MN buildup system was calibrated and its uncertainty was evaluated and verified. The researchresultsshow that the deformation of the buildup structure would induce indication error in the buildup system. The uncertainty component induced by the indication error is larger than that induced by the force transducer,whichis the main component in the combined uncertainty of buildup system with a superlarge range. In the evaluation, the influence of this component should be fully considered; when extrapolating the indication error and its uncertainty to full range, the power function should be used to perform the curve fitting and extrapolation of the buildup system indication error and its uncertainty, which will guarantee thatthe extrapolation results are accurate and reliable.